AMT-Ex Dewpoint Transmitter for Hazardous Applications

  • Certified Intrinsically Safe to IECEx, ATEX for use in hazardous areas and with potentially explosive atmospheres using a specified Signal Isolator in the safe area
  • Ultra-high capacitance aluminium oxide sensor delivers unsurpassed sensitivity, speed of response, accuracy, repeatability, stability, and long service life
  • Unique AutoCal span correction feature
  • Various dewpoint ranges available from -110 °C to 20 °C
  • Compact 3-Wire 4-20 mA Intrinsically Safe Dewpoint Transmitter
  • Accuracy ±2 °C dewpoint
  • IP66 / NEMA 4 Weatherproof Protection
  • Optional Sensor Holder/Sample Cell for optimum sampling conditions
  • Supplied with a calibration certificate traceable to National and International Humidity standards

Model AMT-Ex Dewpoint Transmitter is a 3-wire 4-20 mA powered transmitter that continuously measures moisture in process gas or compressed air. The 4-20 mA linear output signal can be factory configured for any of the following moisture units:- °C or °F dewpoint, ppm(v), ppb(v), g/m3, and lb/MMSCF.

Model AMT-Ex is certified Intrinsically safe to IECEx, and ATEX for use in hazardous areas and with potentially explosive atmospheres using the specified Signal Isolator in the safe area.

At the heart of the AMT-Ex Dewpoint Transmitter is our ultra-high capacitance aluminium oxide sensor, which delivers unsurpassed sensitivity, speed of response, accuracy, repeatability, stability, and long service life. Every unit is supplied with a Calibration Certificate traceable to International Humidity Standards, guaranteeing accuracy to ± 2 ºC dew point.

Read More

Transmitter Type 3-Wire 4-20 mA Intrinsically Safe  Dewpoint Transmitter
Sensing Element Ultra-High Capacitance – Aluminium Oxide Type
Output Signal Linear 4-20 mA with selectable engineering units and fault indication
Power Supply 12-28 Vdc Reverse Polarity Protected
Dewpoint Ranges See datasheet
Intrinsic Safety Certification IECEx BAS07.0080 BASEEFA 07ATEX0234 BAS21UKEX0805
Standards EN 60079-0:2018 EN 60079-11:2012
ATEX, UKCA and IECEx Markings Ex II 1GD Ex ia IIC T4Ga Exia IIIC T200 135 °CDA (-20 °C ≤ Ta ≤ +60 °C)
Electromagnetic Compatibility (EMC) Complies with BS EN IEC 61326-1
Accuracy ± 2 °C dewpoint (UK National Physical Laboratory traceable for range -90 °C to 20 °C)
Repeatability Better than ±0.3 °C dewpoint
Operating Temperature (Ambient and Process) -20 °C to 60 °C
Operating Pressure (Process) From 1 kPa (0.01 bara) to Maximum 35,000 kPa (350 bara)
Operating Humidity (Ambient) Maximum 95 %RH Non-condensing
Storage Temperature -20 °C to 70 °C
Field Calibration AutoCal span check and correction are performed using two buttons on the Transmitter body. Locked in normal operating mode to avoid tamper. We recommend AutoCal is repeated every 2-3 months
Factory Calibration Supplied with a calibration certificate traceable to National and International Humidity standards. We recommend annual laboratory calibration
Sample Flow Rate Flow independent, ideally 5 to 15 L/min, maximum 20 L/min
Cable Terminations IP66 / NEMA 4X rated, size C, DIN EN 175301 connector at the Transmitter and other end terminated with bootlace ferrules
Cable Supplied with 2 m IS (blue) standard cable. Nominal diameter 3.4 mm, 92 ohms/km at 20°C
Mechanical Connection 3/4” UNF (16tpi) with integral Viton “O” ring seal
Isolation Sensing Element connected to the 4-20 mA loop but isolated from Transmitter body
Transmitter Body 316 Stainless Steel
Sensor Protection 316 Sintered Stainless Steel Filter – 50 micron
Probe Material (Wetted Parts) 316 Stainless Steel
Ingress Protection IP66 / NEMA 4X when Connector coupled to Transmitter
Manufacturer’s Warranty 12 months in case of defective parts or faulty workmanship
  • Calibration Certificate traceable to National and International Humidity standards
  • 2 m Connecting Cable

Related Products